AII GPR-1500 PPM Trace Oxygen Analyzer
RM 0.00
Description
Trace Oxygen Analyzer – AII GPR-1500 PPM
Wall Mounted Easy To Maintain PPM Oxygen Measurement
AII GPR-1500 PPM Trace Oxygen Analyzer is a wall mounted oxygen analyzer in a NEMA enclosure, suitable for indoor or outdoor. This single channel oxygen monitor is supplied with a stainless steel O2 sensor housing for trace oxygen gas analysis in safe as well as hazardous areas where ATEX approval. The low detection limit of 0.05ppm allows a minimum oxygen measurement range of 0 to 10 ppm O2. Ideal for clean gas applications and when used with the XLT oxygen sensor type it allows measurement in pure carbon dioxide. The Analytical Industries long-life oxygen sensors are maintenance-free sealed electrochemical cells that don’t need top up of electrolyte and can be exchanged in few minutes.
AII GPR-1500 PPM Trace Oxygen Analyzer Features:
- Measurement ranges 0 to 10ppm up to 0 to 1% O2
- Can be calibrated with ambient air.
- General purpose or intrinsically safe (ATEX) models
- Fast recovery from exposure to air
- Can measure in pure CO2 with XLT sensor
- Two alarms with GPR-1500 A analyzer model
Model & ranges:
GPR-1500A | GPR-1500N | GPR-1500N(ATEX) | |
Measurement range |
0-10, 0-100, 0-1000 ppm, 0-1%, (0-25%) |
||
Accuracy |
< 2% of selected range at constant conditions |
||
Response time |
T90 < 10 seconds |
||
Sensitivity (LDL) |
0.05 ppm |
||
Linearity |
<1% of scale |
||
Sensor model |
GPR-12-333 for inert gas streams |
||
Sensor life at 25ºC (77°F) and 1 atm |
24 months in < 1000 ppm O2 |
||
Calibration interval |
30 days |
||
Inlet pressure |
0.34-2 barg (5-30 psig) with atmospheric vent |
||
Flow rate |
0.5-1.0 l/m (1-2 SCFH) |
||
Gas connections |
1/8″ or 1/4″ compression tube fittings |
||
Wetted parts |
Stainless steel |
||
Display |
Graphical LCD 7 x 3.5cm (2.75 x 1.375″); resolution 0.01 |
||
Enclosure |
Fiberglass NEMA 4X, 6.75 x 8.375 x 4.25″ |
||
Weight |
3.6kg (8lbs) |
||
Compensation |
Temperature compensated output |
Temperature | |
Signal output | 4-20 mA or 1-5 V, optional Modbus |
4-20 mA |
|
Alarms | Two user configurable alarms |
No alarms |
|
Operating temperature | GPR sensor: 5°C to 45°C (41°F to 113°F) XLT sensor: -10° to 45°C (14°F to 113°F) |
GPR sensor: 5°C to 45°C (41°F to 113°F) |
|
Power |
12-28 V DC non-loop or 110-220 VAC |
18-24 V DC | |
Area classification |
General purpose |
UL: Class I, Division 1, Group C, D hazardous areas with external intrinsic safety barrier ATEX: II 2 G Ex ia IIB T4 Tamb -20°C to +50°C |
|
Recovery time |
30 seconds in air to < 10 ppm in < 1 hour on N2 purge |
Additional information
Model | GPR-1500 A, GPR-1500 N, GPR-1500 N ATEX |
---|