AII GPR-1500 GB/2500 GB Oxygen Monitors

RM 0.00

Clear

Description

Oxygen Monitors – AII GPR-1500 GB/2500 GB

O2 Monitors for Glove Boxes and Nitrogen Generators

 

AII GPR-1500 GB/2500 GB Oxygen Monitors designed specifically for measuring oxygen in glove boxes from air (21% O2) down to low levels (0-10 ppm O2), this oxygen transmitter range has easy installation options that ensure the best fit with the glove box operators needs. The analyzer or remote sensors can be connected directly on the glove box using a KF-40 flange. In the remote version the oxygen sensor can be placed in the glove box using our sample/calibration module.

 

AII GPR-1500 GB/2500 GB Oxygen Monitors Features:

  • Measurement ranges from 0-10ppm up to 0-25% O2
  • Accuracy of better than 2% of range
  • 24 months sensor life span (in normal use)
  • 4-20 mA output
  • 18-24V loop powered
  • XLT sensor options for CO2 backgrounds

 

Model

  GPR-1500 GB GPR-2500 GB
Measurement range 0-10, 0-100, 0-1000 ppm, 0-1%,
0-25% (calibration only)
0-1%, 0-5%, 0-10%, 0-25%
Accuracy < 2% of selected range at constant conditions Accuracy
Response time T90 < 10 seconds Response time
Recovery time 60 seconds in air to < 10 ppm in < 1 hour on N2 purge Not applicable
Sensitivity (LDL) 0.05 ppb 0.01%
Linearity < 1% of scale < 1% of scale
Sensor model GPR-12-333-M

XLT-12-333-M for gases containing > 0.5% CO2

GPR-11-32-4

XLT-11-24-4 for gases containing > 0.5% CO2

Sensor life at 25ºC (77°F) and 1 atm 24 months in < 1000 ppm O26 months in air GPR-11-32-4 32 months in air
XLT-11-24-4 24 months in air
Calibration interval 30 days 31 days
Inlet pressure

Nominally atmospheric
Flow through system: 0.34-2 barg (5-30 psig) with atmospheric vent

Flow rate

Ambient monitoring or Flow through system 0.5-1 Nl/min (1-2 SCFH)

Gas connections

KF 40 flange or 1/8” compression tube fittings (with flow through housing)

Wetted parts

GPR-1500 GB Stainless steel GPR-2500 GB Delrin

Display Graphical LCD 7 x 3.5cm (2.75 x 1.375″); resolution 0.001 Graphical LCD 7 x 3.5cm (2.75 x 1.375″); resolution 0.002
Enclosure

Painted aluminum, 7.6 x 10.1 x 5.1cm (3 x4 x2″)

Weight

900g (2lbs)

Compensation

Temperature

Signal output

4-20mA (loop current)

Alarms

None

Operating temperature

GPR sensor: 5°C to 45°C (41°F to 113°F)
XLT sensor: -10°C to 45°C (14°F to 113°F)

Power

18-24 V DC two wire loop

Area classification

General purpose

Additional information

Model

GPR-1500 GB, GPR-2500 GB