AII GPR 1600/2600/3100 Oxygen Analyzers

RM 0.00

Clear

Description

AII GPR 1600/2600/3100 Oxygen Analyzers for Industrial Gas

Percent and PPM O2 Monitoring in Bulk Gases

AII GPR 1600/2600/3100 Oxygen Analyzers range of oxygen analyzers are designed to monitor gas produced or used in industrial processes where trace oxygen from low parts per million to pure O2 must be precisely measured. Utilizing AII’s high performance galvanic oxygen sensors these advanced instruments are simple to use with a common HMI and menu structure across-the-range chassis. Operators only need to learn just one instrument for multiple gas analysis applications at various oxygen levels. Available as bench-, 19″ rack-, panel- and wall-mounted for flexible installation.

 

AII GPR 1600/2600/3100 Oxygen Analyzers Features:

  • High accuracy ( ±0.02 ppm in lowest range)
  • 24 months sensor life with normal usage
  • Sample in CO2 backgrounds with XLT sensor
  • Four measurement ranges with manual or auto-ranging
  • Two Alarms and range indication output
  • Bench, rack or wall mounting options
  • Integrated bypass valve (for GPR-1600)
  • Easy access and sensor replacement
  • Simple to use HMI

 

Model

GPR-1600 GPR-2600 GPR-3100
Measurement range 0-10, 0-100, 0-1000 ppm, 0-1%
(0-25% calibration only)
0-1%, 0-5%, 0-10%, 0-25% 0-100%, plus suppressed zero ranges.
50, 80, or 90-100%
Accuracy

< ±2% of range
±0.1% at constant conditions after calibration with 95-100% oxygen (GPR-3100)

Response time

T90 < 10 seconds

T90 < 13 seconds
Recovery Time 60 sec in air to < 10 ppm in < 1 hour on N2 purge

N/A

Sensitivity (LDL) 50 ppb 50 ppm 0.1% oxygen
Linearity

< 0.5% of scale

Sensor model GPR-12-333

XLT-12-333 for gas mixture with > 0.5% CO2

GPR-11-32-4

XLT-11-24-4 for gas mixture with > 0.5% CO2

GPR-11-120-OP
Sensor life at 25°C and 1 atm 24 months in < 1000 ppm O2 GPR-11-32-4 32 months; XLT-11-24-4 24 months 24 months in 100% oxygen
Calibration interval

Typically: 1-3 months

Inlet pressure

0.34 – 2 barg (5-30 psig) with atmospheric vent

248 bar g (3600psig)
Flow rate

0.5 – 1.0 l/m (1-2 SCFH)

Gas connections

1/4″ compression tube fittings

Display

Graphical LCD 12.7 x 7cm (5 x 2.75″); resolution 0.01

Enclosure

Painted aluminum
See individual mounting options for dimensions

Compensation

Barometric pressure and temperature;
Temperature controlled sample system and sensor (GPR-3100 only)

Analog output

4-20 mA isolated, 0-1V, and 0-5V

Range ID

1-5 V or 4-20mA, optional relay contacts

Communications

Choose from RS485, RS232 or USB

Alarms

Two user-adjustable alarm relays

Operating temperature

GPR sensor: 5°C to 45°C (41°F to 113°F)
XLT sensor: -10°C to 45°C (14°F to 113°F)

Power

Universal 100-240 V AC

Additional information

Model

GPR-1600, GPR-2600, GPR-3100